Electromagnetic Immunity For Non-Life-Supporting Me Equipment And Me Systems - Zeiss S7 / OPMI VISU 160 Instructions For Use Manual

Table of Contents

Advertisement

System Data
Electromagnetic immunity for non-life-supporting ME equipment
and ME systems
Interference re-
IEC 60601
sistance
test level
Tests
Conducted RF dis-
3 V
turbances as de-
150 kHz to 80 MHz
scribed in
EN 61000-4:-6
Radiated RF distur-
3 V
bances as de-
80 MHz to 2.5 GHz
scribed in
EN 61000-4:-3
Note 1
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2
These guidelines may not apply in all situations. Electromagnetic propagation is influ-
enced by absorption and reflection by structures, objects and persons.
Page 188
The device S7 / OPMI VISU 160 is intended to be operated in one of the elec-
tromagnetic environments specified below. The customer or the user of the
device S7 / OPMI VISU 160 is responsible for ensuring that it is operated in
such an environment.
Compliance level Electromagnetic environment -
3V
actual value
3 V/m
actual value
Guidelines
Portable and mobile radio communication
equipment should not be used closer to the de-
vice S7 / OPMI VISU 160, including cables, than
the recommended safety distance that is calcu-
lated using the equation applicable to the trans-
mission frequency involved.
Recommended safety distance:
,
d
1 2 P
=
,
d
1 2 P
for 80 MHz to 800 MHz
=
,
d
2 3 P
for 800 MHz to 2.5 GHz
=
where P is the output power rating of the trans-
mitter in watts (W) according to the transmitter
manufacturer's specifications and d is the rec-
ommended safety distance in meters (m). The
field strength of stationary transmitters as deter-
mined by a site survey should be
compliance level in all frequency ranges.
Interference may occur in the vicinity of equip-
ment marked with the following symbol.
S7 / OPMI VISU 160
a
less than the
b
Version 6.0
G-30-1528-en

Advertisement

Table of Contents
loading

Table of Contents