System Overview - Zeiss S100 / OPMI pico Instructions For Use Manual

Hide thumbs Also See for S100 / OPMI pico:
Table of Contents

Advertisement

Device overview

System overview

Surgical microscope
S100 suspension system
Light source
Page 52
The surgical microscope (4) enables the illumination and visual magnification
of the field of view. For visual magnification, the surgical microscope can
be equipped with different tubes, eyepieces and objective lenses.
The S100 comes as a floor stand, ceiling mount or wall mount. It is also
available as a support unit and for adaptation to a Centro stand.
The S100 is used to control and supply the functions of the surgical
microscope. It consists of a carrier arm (2), suspension arm (3) and supply
unit (1) with light source. The suspension arm allows rough positioning of
the surgical microscope and moving into standby position.
The S100 floor stand can be positioned in the OP by four controllable casters
(5). The locking tabs of the casters prevent the system from being moved or
rolling away unintentionally.
Depending on the order, the device may be equipped with a halogen,
xenon or LED light source. The halogen, xenon or LED light source is used
to illuminate the surgical area. It has a fiber illumination in the surgical
microscope, i. e. the light is directed from the light source via a light guide
to the surgical microscope. The halogen, xenon or LED light source is situated
in the lamp housing of the supply unit (1).
The halogen light source contains two halogen lamps. The second lamp is
used as backup lamp.
The xenon light source produces light whose spectrum is similar to daylight.
Regardless of the brightness setting, the color temperature of the light always
remains the same. Normal daylight film without any additional conversion
filters can therefore be used for photographic documentation.
The xenon light source contains a lamp module with two xenon lamps.
The second lamp is used as backup lamp.
The LED light source has an integrated failure concept. In the event of a defect
in the LED light source, the light intensity drops to 50% and the control lamp
on the lamp housing lights up.
S100 / OPMI pico
Version 9.0
G-30-1781-en