Operation

Checklist

Page 146
For safety reasons, the device must not be used if a malfunction was noticed
before the start of a procedure.
• Correct the malfunction (see "What to do in the event of malfunctions")
or contact ZEISS Service.
 Check that the correct rated voltage has been set on the suspension sys-
tem.
 Check that all cables have been connected.
 Check that the light guide has been connected.
 Check that the OPMI protective cover has been removed.
 There is enough room for the accessories installed.
• Turn on the system at the power switch of the suspension system.
Surgical microscope
Check the zoom
• Press the appropriate button on the foot control panel.
Check the focus
• Press the appropriate button on the foot control panel.
Friction adjustment of the surgical microscope
 Check that the friction of the surgical microscope's rotation has been ad-
justed as required using the friction adjustment knob on the suspension
system.
Speeds of the microscope functions
 Check that the speeds of the microscope's functions have been adjusted
on the S7 suspension system as required.
Eyepieces / binocular tube
 Check that the surgical microscope and the tube are in an ergonomic po-
sition.
 Check that the correct interpupillary distance has been set.
 Check that the eyecups have been adjusted in such a way that you can see
the full field of view.
 Check that the correct prescription has been set on the diopter scale.
 Image quality is correct throughout the entire magnification range.
S7 / OPMI VISU 160
Version 6.0
G-30-1528-en