Preparations for use

Adjusting the system - General

Setting up the suspension system

CAUTION
Page 134
Retinal injury to the patient's eye!!
Too much light intensity (brightness setting is too high) or excessive radiation
exposure times may lead to retinal injury in the patient's eye.
• Adjust the illumination intensity as required for the selected type of illu-
mination and the radiation exposure time. You will find the values recom-
mended by ZEISS in the table "Maximum radiation exposure times" on
Page 21.
• Turn on the unit at power switch (5).
• Use arrow keys (4) to successively select the following functions:
lamp brightness,
motor speeds for focusing, zoom and X-Y coupling.
 The respective LED (2) lights up.
• Use keys (3) to set the values required for:
lamp brightness,
motor speeds for focusing, zoom and X-Y coupling.
 The respective setting can be read from the display (1).
Lamp brightness
setting range: 0.1 to 1.5
Level 0.1 corresponds to the lowest, level 1.5 to the highest lamp bright-
ness.
Motor speed
Setting range: 1 to 10
Level 1 corresponds to the lowest, level 10 to the highest motor speed.
– The value set is saved after 10 seconds and displayed the next time the
system is started.
– If no entry is made via the arrow or plus/minus keys, the system will switch
to the lamp brightness display after 15 seconds at the latest.
S7 / OPMI VISU 160
Version 6.0
G-30-1528-en