Safety check

CAUTION
Version 6.0
G-30-1528-en
Risk of injury!
Cables and plug connections may come loose and parts of the housing may
become live.
• Make sure that the regular technical safety checks required for this sys-
tem in accordance with the applicable national regulations are performed
on schedule and to the stipulated extent.
The safety checks must only be performed by the provider of your overall
system.
The scope of the safety checks of the system should comprise at least the fol-
lowing points:
Availability of the Instructions for Use
Visual inspection of the system and accessories for damage and legibility
of the labels
Leakage current test
Test of protective ground conductor
Function and wear test of the steerable casters and locking tabs
Function test of all switches, buttons, sockets and indicator lamps of the
system