Connecting The Suspension System - S7 Floor Stand - Zeiss S7 / OPMI VISU 160 Instructions For Use Manual

Table of Contents

Advertisement

Preparations for use

Connecting the suspension system - S7 floor stand

NOTE
CAUTION
Page 118
Risk of tripping!
Inappropriate routing of cables increases the risk of tripping.
• Always run cables in a manner that will not impede the user's move-
ments.
• Check the voltage indicator at (5).
The suspension system is set at the factory to the rated voltage used in the
country of destination. The rated voltage indicated in window (5) must
correspond to the rated voltage available on the site of installation.
• If the rated voltage has not been correctly set, adjust the sliding switch (5)
using a suitable tool.
Insert or remove connectors at (2) and (3) only when the power switch is off.
• Plug the connector of the foot control panel into connector (2) on the sus-
pension system if you wish to use a foot control panel.
Electrical voltage!
Risk of electrical shock
• Only connect the components to wall outlets which are provided with a
properly connected protective ground conductor.
• Connect the stand to line power by plugging the power cord supplied for
this purpose into (4).
• Connect the suspension system with the potential equalization panel in
the OR, if required. Use the potential equalization pin (1) for this purpose.
S7 / OPMI VISU 160
Version 6.0
G-30-1528-en