Carrier And Suspension Arms - Zeiss S7 / OPMI VISU 160 Instructions For Use Manual

Table of Contents

Advertisement

System Overview

Carrier and suspension arms

-
+
Balance
Page 88
1 Lamp housing of the halogen light source
The lamp module contains a backup lamp which is automatically swung
into the illumination beam path when the first lamp fails.
2 Friction
Knob for adjusting the friction of the suspension arm's upward/downward
movement.
3 Locking lever for limiting downward movement
You can limit the downward movement of the suspension arm by adjust-
ing a stop.
Setting range: from the horizontal position of the suspension arm to its
lower stop.
4 Balance setting
Knob for setting the spring force for balancing. After mounting the sur-
gical microscope including all accessories, adjust the balance setting of the
suspension arm using this screw. The balance setting procedure is de-
scribed in the chapter "Preparations for use".
5 Friction
Knob for adjusting the friction of the suspension arm's swivel movement.
6 Friction
Knob for adjusting the friction of the carrier arm's swivel movement.
7 Control and display panel
The control and display panel permits you to control all electrical functions
of the S7 suspension system and the surgical microscope.
8 Connector panel
Further connectors are provided for powering peripheral equipment. A
foot control panel (option) can be connected to a multi-point connector.
S7 / OPMI VISU 160
Version 6.0
G-30-1528-en