System Overview
Light source
NOTE
Page 68
The illumination system consists of a halogen light source and has been de-
signed for fiber illumination. The lamp housing contains a backup lamp which
automatically swings into the illumination beam path when the first lamp fails.
1 Lamp module
2 Ventilation grid
Overheating of the lamp module!
If ventilation grids are covered e.g. by drapes, this may lead to overheating
of the lamp housing and deactivation of the lamp.
• Never cover the ventilation grids.
• Lamps switched off due to overheating will be automatically switched on
again when they have cooled down.
3 Flap
The flap is the mechanical indicator for the operating status of the halogen
lamps..
–
When the flap is closed, the main lamp is operative.
–
When the flap is open, the main lamp has failed. The backup lamp is
on.
4 Manual activation of the backup lamp
If the automatic activation function fails, press this button to switch on the
backup lamp.
5 Opening the lamp module
When you press this button, the lamp module is partially ejected. Pull out
the lamp module all the way to replace the bulb.