Light Source - Zeiss S7 / OPMI VISU 160 Instructions For Use Manual

Table of Contents

Advertisement

System Overview

Light source

NOTE
Page 68
The illumination system consists of a halogen light source and has been de-
signed for fiber illumination. The lamp housing contains a backup lamp which
automatically swings into the illumination beam path when the first lamp fails.
1 Lamp module
2 Ventilation grid
Overheating of the lamp module!
If ventilation grids are covered e.g. by drapes, this may lead to overheating
of the lamp housing and deactivation of the lamp.
• Never cover the ventilation grids.
• Lamps switched off due to overheating will be automatically switched on
again when they have cooled down.
3 Flap
The flap is the mechanical indicator for the operating status of the halogen
lamps..
When the flap is closed, the main lamp is operative.
When the flap is open, the main lamp has failed. The backup lamp is
on.
4 Manual activation of the backup lamp
If the automatic activation function fails, press this button to switch on the
backup lamp.
5 Opening the lamp module
When you press this button, the lamp module is partially ejected. Pull out
the lamp module all the way to replace the bulb.