System Combinations - Zeiss KSK 150 FC Instructions For Use Manual

Table of Contents

Advertisement

KSK 150 FC

System combinations

System kit
Version 7.1
G-30-1158-en
The Colposcope KSK 150 FC can be extended with the accessories offered by
Carl Zeiss Meditec A (see ordering data for accessories) to form an already
tested system. Then this system combination is approved by ZEISS.
It has to be noted that all devices used in the system have to be cleaned in
accordance with their relevant instruction for use.
Any additional equipment connected to medical electrical devices must de-
monstrably comply with the applicable standards and directives (e.g. IEC
60950-1:2005, modified + Cor.:2006 + A1:2009, modified + A1:2009/
Cor.:2012 + A2:2013, modified for data processing equipment).
Furthermore, any configurations must comply with normative requirements
for electronic medical devices (see IEC 60601-1-1:2000 or IEC 60601-1:2005
+ A1:2012, chapter 16).
Anyone connecting additional equipment to medical electrical devices is a
system configurer and as such is responsible for compliance of the system
with the normative requirements for medical electrical systems.
Please note that local legislation takes priority over the above-mentioned nor-
mative requirements. If you have any questions, please contact your local
dealer or ZEISS Service.
• Never connect any multiple connectors.