Service and warranty concept
Comparison of fixed and condition-based service interval
Event
Maintenance
The maintenance intervals that are actually used depend on the
country/region the CPO is located.
Concept with fixed
service interval
Knee joint is maintained
every 24 months
Service and warranty concept | Kenevo 3C60 | Prosthetic | Global Academy | 22.11.2020
Condition-based
service interval
Knee joint is maintained
when defined condition
parameters have been
reached (or latest after 36
months)
| © Ottobock