User Instructions
Safety modes – Drained battery
< 5%
In the case of a drained battery, the prosthesis switches
back to basic mode after the battery is charged!
•
This safety mode is displayed with 10 long beep and
vibration signals (Battery level < 5%)
•
Stance flexion resistance corresponds to the 'Activity Mode':
•
Mode A / B / B+:
•
Knee joint with very high flexion resistance
→
Allows safe standing and to flex the knee joint
during sitting down
•
Mode C:
•
Adjusted stance flexion resistance
•
No stance release possible