Alignment and adjustment
Static alignment optimization with L.A.S.A.R posture
2
Weight line
(Measurement with both feet
1
Measurement with prosthetic foot
)
≤ 15 mm
Alignment and adjustment | Kenevo 3C60 | Prothetik | Global Academy | 12.12.2016
L.A.S.A.R posture provides also a
method to check if significant postural
compensations present
Determine difference of both
measurements:
Load line (measurement only with
prosthetic foot):
Load line
AND
Weight line (measurement with
both feet)
If variation is ≤ 15 mm = OK (only slight
compensation present)
3 cm in front of knee axis
(important for knee safety)