Configuring; Resetting To Default Settings - wtw Cond 3150i Operating Manual

Conductivity pocket meter
Table of Contents

Advertisement

Available languages

Available languages

Configuring

t
2 5
Tref25
S/
o
d
C
n
Basic
settings
I
n
I
8
Configuring
Note
You can leave the configuration menu at any time with <M> . The parameters
that have already been changed are stored.
Procedure for configuring (factory settings appear in bold typeface):
Switch off the measuring instrument.
While pressing the <M> key, press <
Switch over the reference temperature of the conductivity with <CAL>.
Selection: 25 °C (Tref25) and 20 °C (Tref20).
Then press the <RUN/ENTER> key.
Switch over the measurement value display with <CAL>.
Selection: Conductivity (S/cm) or specific resistance (MΩ
cm
Press the <RUN/ENTER> key. The instrument changes to the measur-
ing mode.

Resetting to default settings

The following settings are reset to the delivery status when a reset is per-
formed (initialized):
Measuring mode
Cell constant
Temperature compensation
Reference temperature
Reset procedure:
Press <CAL> while pressing the <RUN/ENTER> key.
Press <RUN/ENTER> to reset the measurement parameters. The instru-
ment changes to the measuring mode.
>.
-1
0.475 cm
Nonlinear (nLF)
25 °C (Tref25)
Cond 3150i
cm).