Measuring The Temperature - wtw Oxi 7310 Operating Manual

Compact precision dissolved oxygen meter
Table of Contents

Advertisement

Dissolved oxygen
24
Measured
Time interval
parameter
D.O. concen-
10 seconds
tration
D.O. satura-
10 seconds
tion
D.O. partial
10 seconds
pressure
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.

5.1.2 Measuring the temperature

For reproducible D.O. measurements, it is essential to measure the tempera-
ture of the test sample.
The temperature is automatically measured by the temperature sensor
(NTC30 or Pt1000) integrated in the sensor.
The CellOx 325 / StirrOx
integrated temperature sensor.
Stability in the time inter-
val
Δ : better than 0.05 mg/l
Δ : better than 0.6 %
Δ : Better than 1.2 mbar
®
G and DurOx 325 D.O. sensors have an
Oxi 7310
ba75931e03
12/2011