5.1.2 Measuring the temperature

For reproducible pH and ORP measurements, it is essential to measure the
temperature of the test sample.
You have the following options to measure the temperature:
 Automatic measurement of the temperature with the temperature sensor
(NTC 30 or Pt 1000) integrated in the sensor.
 Measurement with an external temperature sensor.
 Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
Temperature
Resolution of the
sensor
yes
0.1 °C
-
1 °C
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
1.
Measure the current temperature of the test sample.
2.
Set the temperature value with <><>.
or
In the <F1>/[Menu]/Man. temperature menu, set the temperature value
with <><>.
temp. display