Equipment Classification

Per EN/IEC60601-1, Medical Electrical Equipment, General Requirements for
Safety, this device is classified as following:
Table 6: Equipment classification
Class II equip-
ment
Ingress
Cleaning
Disinfection
Flammable
anesthetics
Operation
Related Links
Cleaning and Disinfecting
3253F EN 20210312 1125
Equipment in which protection against electric shock does
not rely on basic insulation only, but in which additional
safety precautions such as double insulation or reinforced
insulation are provided, there being no provision for protec-
tive earthing or reliance upon installation conditions.
This device does not have protection against ingress of wa-
ter.
See section on cleaning and disinfecting.
See section on cleaning and disinfecting.
This device is not suitable for use in the presence of a flam-
mable anesthetic mixture with air, or in presence of a flam-
mable anesthetic mixture with oxygen or nitrous oxide.
Continuous operation.