Outlet Connections; Pump Inlet Connections - Teledyne SyriXus 65x Installation And Operation Manual

Syringe pumps
Table of Contents

Advertisement

SyriXus Series Pumps Installation and Operation Guide
Section 4 Gradient Pumping for Pressure, Flow, and Concentration Modes
Figure 4-2 Pump inlet connections

4.2.2 Outlet Connections

EAR99 Technology Subject to Restrictions Contained on the Cover Page
4-4
FROM REFILL
KIT ASSEMBLY
spacer block
2-way
valve
Valve components and connections are shown in Figure 4-3.
1. Mount the two-way outlet valve on the side of the pump housing
opposite the refill valve, with the spacer block and screws pro-
vided.
2. Connect the pre-bent SST tubing between one port of the valve
and the pump outlet. Use the nut and ferrule to connect the tub-
ing at the outlet and the valve fittings to connect the tubing at
the valve.
3. Connect the 5.1 cm length of 1/8" tubing to the other port of the
valve, using the valve fittings.
4. Connect the 1/8" side of the reducing union to the tubing.
5. Connect the 1/16" side of the reducing union to the 1.5 m length
of 1/16" tubing. This tubing may be cut to an appropriate length.
pre-bent
SST tubing
SST tubing
TO
APPARATUS