SyriXus Series Pumps Installation and Operation Guide
Section 8 Serial Interface
Command (Refer to
NOTE 1 & NOTE 2
IDENTIFY
INDEPENDENT
INDEPENDENTCD
IPUMPA=1, IPUMPA=0
IPUMPB=1, IPUMPB=0
IPUMPC=1, IPUMPC=0
IPUMPD=1, IPUMPD=0
LGE,F:XX,A:0X
LGGO
LGSL,F:xx
LGST
LGDL,F:xx,S:xx
LGUL,F:xx,S:xx
EAR99 Technology Subject to Restrictions Contained on the Cover Page
8-22
Table 8-3 Serial Commands (Continued)
.)
Pump responds 'SERIES=1240-0___, MODEL ____D PUMP; REV__.' For each
pump, REV__ is the internal pump program software revision. (For example, if the
controller was attached to two 100DMs, the message would read
'SERIES=1240-024, MODEL 100DM PUMP; SERIES=1240-024, MODEL 100DM;
REV__.') The series number is the original catalog number for the pump type. It
may not match the production series number on the pump serial label.
SERIES=1240-024, MODEL 100DM PUMP
SERIES=1240-027, MODEL 100DX PUMP
SERIES=1240-021, MODEL 260D PUMP
SERIES=1240-025, MODEL 500D PUMP
SERIES=1240-103, MODEL 500HPx PUMP
SERIES=1240-052, MODEL 1000D PUMP
SERIES=1240-063, MODEL 65D PUMP
SERIES=1240-096, MODEL 65DM PUMP
SERIES=1240-819, MODEL 30D PUMP
Put pumps in Independent Mode.
Turns the pressure integral control On and Off for the pump indicated.
1 = ON
0 = OFF
Action to perform when gradient program reaches the end. The pump needs to be
in local mode only.
File # F: 01-99
Action A:
00=Hold final value
01=Stop after final step
02=Return to initial value and hold
03=Return to initial value and repeat program
Start Gradient Command. This starts a gradient program (same as the 'RUN' key).
This command will check to see if there is a gradient running and respond with
'RUNNING' if there is. The pump needs to be in local mode only.
Select Gradient File Command. This selects a gradient file to be run. This com-
mand will reset the controller to the saved file gradient type. If the selected gradient
file does not exist, the controller will respond with 'PROBLEM=INVALID OPER-
AND.' The pump needs to be in local mode only.
Stop Gradient Command. This stops a gradient program. The pump needs to be in
local mode only.
Gradient Step Download command. This downloads a step from the pump to the
PC. This command will respond with 'PROBLEM=INVALID OPERAND' if the file or
step does not exist. The controller will respond with step information if the com-
mand is valid. Refer to Section 8.8.2 "Gradient Download Command" for complete
information on Gradient step download commands. The pump needs to be in local
mode only.
Gradient Step Upload command. This transfers a step from the PC to the controller.
Refer to Section 8.8.3 "Gradient Upload Commands" for complete information on
Gradient step download commands. The pump needs to be in local mode only.
Description