SyriXus Series Pumps Installation and Operation Guide
Section 9 Pump Maintenance, Troubleshooting, and Servicing
9.5 Resetting the
System

9.5.1 Basic Reset

9.5.2 Hard Reset

9.6 Diagnostic Menu

Figure 9-1 Accessing the diagnostic menu

9.6.1 LCD TEST

EAR99 Technology Subject to Restrictions Contained on the Cover Page
9-4
This action completely deletes user programmed settings, erasing
all programs and returning the controller to factory default settings.
Record your program settings and parameters before performing a
reset operation. If
ZERO PRESS
will be lost.
To perform a basic reset, press
(A).
To perform a hard reset:
1. Turn the controller switch to
2. Press and hold
CLEAR ENTRY
3. While still holding
Continue holding down
4. Release the
CLEAR ENTRY
to
.
STANDBY
5. Turn the controller switch back to ON, and perform the Basic
Reset, as described above.
The syringe pump controller can run a number of system tests that
are initialized using the keypad.
1. LCD TEST
2. ANALOG INPUTS
3. A/D CAL
MORE
Cycles all segments of the display through all displayed characters. To
stop the test, press any key.
has been employed, all corrected offsets
>
MENU
SYSTEM RESET
.
STANDBY
.
, turn the controller switch to
CLEAR ENTRY
for one second.
CLEAR ENTRY
key and turn the controller switch back
(
)
(
MORE
MORE
(
DIAGNOSTIC MENU
4. KEYPAD TEST
5. MOTOR CONTROL/LIMITS
6. SERIAL TEST
(5) >
CONTINUE
.
ON
)
)
PREVIOUS