Shear Key; Accessing The Shear Key; Replacement Shear Keys - Teledyne SyriXus 65x Installation And Operation Manual

Syringe pumps
Table of Contents

Advertisement

SyriXus Series Pumps Installation and Operation Guide
Section 9 Pump Maintenance, Troubleshooting, and Servicing

9.13.1 Shear Key

Figure 9-10 Accessing the shear key
EAR99 Technology Subject to Restrictions Contained on the Cover Page
9-18
The shear key is NOT a safety feature! It provides a measure of
protection against damage to the equipment. Avoid
overpressure situations by properly installing and programming
the pumping system. In applications where overpressure
conditions are possible, install an appropriate pressure relief
safety device in your apparatus.
Table 9-1 "Replacement Shear Keys" below provides part numbers for
each pump model replacement shear key.
Table 9-1 Replacement Shear Keys
Pump Model
1000x, 500x, 260x
1. Place the pump on its side and remove the four panhead screws
that attach the access plate to the case bottom, as shown in Fig-
ure 9-10.
2. Remove the cotter pin that passes through the castle nut and
ball screw.
3. Use a 3/4" wrench to remove the castle nut, shown below.
4. Remove the spacer.
5. The two broken halves of the shear key should be protruding
from the brass worm gear and the ball screw.
Access Plate
WARNING
65x
Spacer
Cotter Pin
Part Number
60-1243-607
69-1263-365
Ball Screw
Castle Nut