SyriXus Series Pumps Installation and Operation Guide
Section 4 Gradient Pumping for Pressure, Flow, and Concentration Modes
4.2 Connecting the
System
EAR99 Technology Subject to Restrictions Contained on the Cover Page
4-2
Table 4-1 Manual Refill Valve Kits
Pump Model
1000x
500x
260x
65x
Table 4-2 Manual Outlet Valve Kits
Pump Model
1000x
500x
260x
65x
The syringe pump has two ports at the top of the cylinder. One port is
used as the inlet for filling the pump, and the other as the outlet
(either port may be used as inlet or outlet). Inlet and outlet connec-
tions to each pump must be made identically. Standard plumbing con-
nections vary between pump models. Refer to Table 4-3 "Swaging
Detail" for standard port information.
When making fluid connections that use ferrules, be sure to use the
ferrules supplied for that pump by Teledyne ISCO. Push the tubing
completely into the connector and finger-tighten. Then tighten with a
wrench to clamp the ferrules onto the tubing.
Table 4-3 Swaging Detail
260x
1/8" Valco
500x
1/8" NPT
1000x
1/4" NPT
65x
1/4" F250C
SyriXus series 260x and 65x have a direct connection, as shown in
Figure 4-2 and Figure 4-3. Valve kits for other models include male
adapter fittings.
Part Number
68-1247-117
68-1247-083
68-1247-077
68-1247-127
Part Number
68-1247-118
68-1247-082
68-1247-078
68-1247-126
FERRULE
FRONT FERRULE
BACK FERRULE
NUT 7-16 - 20
FRONT FERRULE
BACK FERRULE
NOTE
NUT 5-16 - 24
1/2 TURN PAST
FINGER-TIGHT
NUT 5-16 - 24
3/4 TURN PAST
FINGER-TIGHT
1-1/4 TURN PAST
FINGER-TIGHT