SyriXus Series Pumps Installation and Operation Guide
Section 8 Serial Interface
Command (Refer to
NOTE 1 & NOTE 2
ZEROA
ZEROB
ZEROC
ZEROD
ZERODIFF1
ZERODIFF2
ZERODIFF3
NOTE 1
The analog input range is -1.5 to 11.6 volts. There is NO conversion of the returned number. The number returned (0
to 65535 decimal) will have an offset of 7500 added to the number (7500 = 0 volts) and a scale of 5000 for every 1
volt, for example:
NOTE 2
The only pump B commands accepted in continuous pumping mode or Modifier Addition Mode are: %B, FLOWB,
LIMITSB, PRESSB, REFILLB, REFILLB=, STATUSB, VOLB.
8.8.1 Get Status
Command
R046=37510,-00128, 01017, 06981, 05743, 00421, 00442, 00449,11111000,CS
Pressure Pump A
Pressure Pump B
Pressure Pump C
Units = (psi*5)
Resolution = 0.2 psi
Figure 8-3 Get status string 'G' command
EAR99 Technology Subject to Restrictions Contained on the Cover Page
8-26
Table 8-3 Serial Commands (Continued)
.)
'Zeros' the pressure sensor offset.
'Zeros' the pressure sensor offset for the respective analog input.
number
7500
= volts
5000
The 'G' and 'G&' serial commands retrieve information from the pump
controller. Each command returns a text string which can be read as
shown in Figure 8-3 and Figure 8-5.
Description
(32500
7500)
= 5 volts
5000
ALOG1 — ALOG5
(refer to Table 8-3)
Units = (volts * 5000)
Resolution = 0.2 millivolt
Range = 7500 to 58035
(1.5000V to 11.6070V)
Digital Inputs 1—8
1 = High
0 = Low