9.6.2 ANALOG INPUTS

Figure 9-2 Analog input diagnostic screen

9.6.3 KEYPAD TEST

9.6.4 MOTOR
CONTROL/LIMITS
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Section 9 Pump Maintenance, Troubleshooting, and Servicing
Displays the values of all analog inputs in volts. In the left column,
with all connected pumps at zero pressure, the pressure readings
should be near zero. Readings for pumps not connected are mean-
ingless.
After the initial reading, pressurize the pump and then observe the
test screen once more to verify that the value for that pump has
increased.
The right column represents the four analog input terminals on the
rear panel of the controller. For any terminal shorted to ground, the
value should read a steady zero.
MORE
Press
(D) to return to the diagnostic menu.
EXIT
The screen will briefly display the name of each key pressed. Press
(D) to return to the diagnostic menu.
EXIT
Tests the digital position controls.
From the Limits menu, you can set minimum and maximum flow rate
limits and pressure limits, as well as the maximum flow rate in Con-
stant Pressure Mode. High and low limits cannot exceed pump specifi-
cations.
The pump stays within these limits by means of a flag that moves up
and down with the ball nut assembly, and two optical sensors at the
top and bottom of the tower side plate as shown in Figure 9-3.
When a sensor is interrupted, it has a logic 1; when it is uninter-
rupted, it has a logic 0. Therefore:
• Upper limit 0, Lower limit 1
• Upper limit 1, Lower limit 0
• Upper limit 0, Lower limit 0 = Cylinder is partially full
• Upper limit 1, Lower limit 1= Illegal state
An illegal state can indicate failure of one or both sensors. The con-
troller display will toggle immediately between
when you press
CYLINDER EMPTY
run).
A
= XX.XXXX
ACC1 = XX.XXXX
B
= XX.XXXX
ACC2 = XX.XXXX
C
= XX.XXXX
ACC3 = XX.XXXX
D
= XX.XXXX
ACC4 = XX.XXXX EXIT
Gnd
or
REFILL
and
CYLINDER FULL
(the pump will not
RUN
9-5