9.7.2 Pump Case Top
Removal
Figure 9-6 Pump case top screws (3 of 6 shown) - TBD
9.8 Test Points - TBD
9.8.1 Power Supply (A1)
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Section 9 Pump Maintenance, Troubleshooting, and Servicing
Some maintenance and troubleshooting procedures require accessing
the pump module interior. Remove the six screws holding the case top
in place (three screws on each side). Lift the cover straight up and off.
The following sections contain tables listing the most commonly used
test points and their voltages. Refer to the controller schematic and
the pump schematic (available at www.teledyneisco.com).
RISK OF ELECTRIC SHOCK - DISCONNECT THE ELEC-
TRIC POWER BEFORE SERVICING. ONLY TRAINED SER-
VICE PERSONNEL MAY REMOVE THE CASE TOP.
RISQUE DE CHOC ÉLECTRIQUE. COUPER L´ALIMENTA-
TION AVANT LA RÉPARATION. L´USAGER NE DOIT PAS
DÉMONTER L´INSTRUMENT OU DÉRANGER LE
MÉCANISME DEDANS. ADRESSER LA REPARATION
SEULEMENT AUX TECHNICIENS COMPÉTENTS.
Output voltages of the power supply can be verified. The voltages in
the pump unit, between circuit common (TP106) and the fuses or at
the test points should be:
F104
F103
F102
TP111
TP110
TP109
The voltage between motor common (TP101) and +VM (TP105) should be +60
to +90 V.
22 VDC to 31 VDC
8.1 VDC to 13 VDC
22 VDC to 31 VDC
15 VDC 0.6 VDC
15 VDC 0.6 VDC
5 VDC 0.3 VDC
9-9