Serial Test; Serial Connection To A Computer - Teledyne SyriXus 65x Installation And Operation Manual

Syringe pumps
Table of Contents

Advertisement

9.6.5 SERIAL TEST

Figure 9-4 Serial connection to a computer
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Section 9 Pump Maintenance, Troubleshooting, and Servicing
Tests the serial channel.
First, confirm operation of the internal port. On the RS-232-C port
on the rear of the controller, short pins 2 and 3 and then run the test.
The screen should display:
Rec: ****** Serial Test #### *****
Sent: ****** Serial Test #### *****
Press any key to exit
The four digits in the R
S
line, with a very slight delay.
ENT
If the unit passes the internal port test, next perform the I/O test. Use
a null-modem cable (available from Teledyne ISCO, P/N 480-7996-00
or most computer parts stores) to connect the 9-pin serial output port
of a computer to the 25-pin serial RS-232-C port, as shown in Figure
9-4.
Avoid using a USB converter cable; test results have shown this not
to be dependable.
Open a terminal emulator program such as HyperTerminal™ (©Hil-
graeve, Inc. Corp. 1287 N. Telegraph Rd. Monroe, Michigan 48162).
The factory default port settings are:
• BPS - 9600. Baud rate should match controller setting (refer to
Section 8.2 "Network Control and Communication")
• Data bits - 8
• Parity - None
• Stop bits - 1
• Flow control - None
line should match the four digits in the
EC
NOTE
9-7