Single-Pump Gradient Programming - Teledyne SyriXus 65x Installation And Operation Manual

Syringe pumps
Table of Contents

Advertisement

4.3 Single-Pump
Gradient
Programming
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Section 4 Gradient Pumping for Pressure, Flow, and Concentration Modes
A single-pump gradient program is based on either pressure or flow,
and controlled by either time duration (in minutes) or rate of change
(units per minute).
The controller's memory can contain a total of up to 200 steps. One
program can contain from 1 to 200 steps.
When operating in Gradient Mode, any connected pumps not used for
gradient are inoperable.
1. To access the gradient programming menus, press PRGM GRAD,
select PRESSURE (2) or FLOW (3), and CONTINUE (C).
The home screen will appear, with either PG (Pressure
Gradient) or FG (Flow Gradient) in the upper left corner.
2. Press PROGRAM (A).
3. At the prompt, enter a file number between 1 and 99. This is the
file name of your program, and can be the name of a new pro-
gram you are creating, or a stored program you want to edit or
run. Press ENTER.
If a selected stored program is in a different mode than that of the
controller selected in Step 1, a brief notification will appear, display-
ing the controller's mode and the file's mode. If you attempt to run
the program without changing the mode of either it or the controller,
the program will not run, and the same message will be displayed.
4. To enter the flow rate for this program, either press
FLOWRATE (A) and use the number keys and Enter, or for
maximum possible flow, simply press MAX (C).
5. To proceed to the programming screen, press STEP FWD (B).
The programming screen will appear, with the file number and
step number at the top of the screen.
PGa
0.000mL/MIN
2PSI
STOPPED
FILE:1
/MIN
FINAL:
PROGRAM
EDIT
REVIEW
NOTE
ENTER FLOWRATE
0.00000
FLOWRATE
PREVIOUS
013.28mL
00:00:00
PSI
OPTION
XX.00000
MAX
4-7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Syrixus 260xSyrixus 500xSyrixus 500xvSyrixus 1000x