Teledyne SyriXus 65x Installation And Operation Manual page 161

Syringe pumps
Table of Contents

Advertisement

The same sequence would occur with all messages. Since the
master and the network controller are combined in this
example, the relaying of messages is not necessary.
Unit 6
R027SERIES=1240-02__, Model ___D PUMP, REV __XX[CR]
The pump responds with identity and software revision letter.
(In this example 02__ would be 021; Model ___D would be 260D;
REV __ signifies the software revision, XX would be replaced by
the correct checksum, which is B4.)
Network Controller and Master Unit #1
6R006REMOTE16[CR]
This places the pump in the Remote Mode.
Unit 6
R 8E[CR]
The pump acknowledges that it accepted the command.
Network Controller and Master Unit #1
6R00ACONST FLOWF8[CR]
This puts the pump into Constant Flow Rate Mode.
Unit 6
R 8E[CR]
The pump verifies that it received the message.
Network Controller and Master Unit #1
6R009FLOW=1.00AB[CR]
This sets the pump's flow rate to 1.00 mL per minute.
Unit 6
R 8E[CR]
The pump verifies that it received the message.
Network Controller and Master Unit #1
6R 58[CR]
Polls the pump.
Unit 1
R 8E[CR]
Pump responds.
Network Controller and Master Unit #1
6R003RUNF0[CR]
The pump is started.
Unit 1
R 8E[CR]
The pump responds.
The system is now running and the network controller continues the
polling scheme. If the controller gives an improper command, the
units will respond with a problem message indicating the type of error.
The format of the message is given in Section 8.8 "Serial Commands
for the SyriXus Series Pumps" and specifies the commands used for
this instrument.
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Section 8 Serial Interface
8-19

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Syrixus 260xSyrixus 500xSyrixus 500xvSyrixus 1000x

Table of Contents