SyriXus Series Pumps Installation and Operation Guide
Section 6 Modifier Addition
6.1.1 Example of
Hardware
6.1.2 Two Independent
Pumps
Figure 6-1 Modifier addition components installation (Two-pump
configuration)
EAR99 Technology Subject to Restrictions Contained on the Cover Page
6-2
The first step is to collect the required hardware. This hardware would
include inline check valves rated for the maximum pressure of the
pumps, a mixing device (possibly a simple fluid tee), along with tubing
and fittings to connect the components together.
This method involves two SyriXus
series pumps (preferably of the
same model) and a controller.
1. Attach one manual refill kit
(available from Teledyne
ISCO) to each of the pumps
as shown.
2. Attach one manual outlet kit
to each of the two pumps
(also available from Teledyne
ISCO).
3. Using metal tubing, attach a
user supplied check valve to
each of the outlet ports on the
outlet valves as shown.
Observe the flow direction
marked on the check valves.
4. Connect the check valves
using metal tubing to the
mixing device (a mixing tee is
shown).
5. Connect the outlet of the mix-
ing device to your system
extraction device.
6. Using appropriate tubing,
connect the refill valve inlet
to the fluid reservoirs.
7. Follow the procedures in Sec-
tion Section 1.5 "Electrical
Connections" to connect the
controller to the pumps.
8. Perform a preliminary check-
out of the pumps as described
in Section 1.6 "Preliminary
Checkout".
9. Fill the pumps with fluid.