Current Loop Output; Output Range Selection Jumpers; Analog Output Options (Connections To Female 25 Pin Sub-D) - Teledyne SyriXus 65x Installation And Operation Manual

Syringe pumps
Table of Contents

Advertisement

SyriXus Series Pumps Installation and Operation Guide
Section 3 Basic Programming and Operation
Figure 3-15 Output range selection jumpers
Table 3-4 Analog Output Options (Connections to Female 25 Pin Sub-D)
Pin No.
Description
1
Chassis (earth)
2
Analog common
3
Analog common
4
Analog common
5
Analog common
14
Output #1
15
Output #2
16
Output #3
17
Output #4
3.14.2 Current Loop
Output
EAR99 Technology Subject to Restrictions Contained on the Cover Page
3-26
A controller running one to two pumps can also output cylinder
volume on the remaining outputs. The volume output range is not
adjustable. Full scale is equal to one pump stroke.
1 or 2 Pump Function
Flow Rate A
Flow Rate B
Volume A
Volume B
For additional information about analog output options, refer to Tech-
nical Bulletin
TB34 Syringe Pump Analog Output
The pump controller can be configured to provide a 4-20mA current
loop output. The 4-20mA analog output board is pre-installed at the
factory. The board initially contains three output cards, for flow rate,
pressure, and volume for Pump A. Additional cards can be factory
installed (up to a total of 12) when the controller is initially purchased.
Refer to Table 3-5 "Analog Output Signal/Sub-D Pins for 4-20mA" for
connections for the 4-20mA and 0-10v 12 channel outputs.
Connections to the controller DB-25 are the same as shown in
Figure 3-14. Refer to Table 3-5 "Analog Output Signal/Sub-D Pins
for 4-20mA" and Table 3-6 "Analog Output Signal/Sub-D Pins for
0-10 VDC" for cable connections to your equipment.
3 Pump Function
Flow Rate A
Flow Rate B
Flow Rate C
Volume B
Options.
NOTE