3.10.1 Constant Flow

3.10.2 Constant Pressure

EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
To set Constant Flow operation, use the following procedure:
1. Press CONST FLOW. "CFa" will be displayed in the upper left
corner of the screen. This denotes that you will be defining con-
stant flow parameters for pump A. If you wish to define parame-
ters for pump B, C, or D, press softkey D, select pump, and then
press softkey 'A', 'B', 'C', or 'D' to select pump A, B, C, or D respec-
tively.
If the main menu is displayed, you must press 'D' under CONST
FLOW.
2. Press 'A' to change the flow rate. The words ENTER FLOW
RATE will flash on the screen.
3. Use the number keys to enter the desired flow rate.
If you make an error, press CLEAR ENTRY to delete one character
at a time.
4. Press ENTER once the desired flow rate is displayed.
5. Press RUN to begin pump operation.
Programming a Constant Pressure operation only requires a few key-
strokes. Use the following procedure:
1. Press CONST PRESS; CPa will be displayed in the upper left
corner of the screen. This denotes that you will be defining Con-
stant Pressure parameters for pump A. If you wish to define
parameters for pump B, C, or D press softkey 'D', select pump,
and then press softkey 'A', 'B', 'C', or 'D' to select pump A, B, C, or
D respectively.
If the main menu is displayed, you must press softkey D under
RETURN before pressing CONST PRESS.
2. Press the softkey 'A' to indicate to the program that you wish to
enter the pressure. The words "ENTER PRESSURE" will flash
on the screen.
3. Use the number keys to enter the desired pressure.