2.4.3 Using the Wrench
Package
Figure 2-9 Secure screws with the 1/4" hex wrench
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Section 2 Fluid System Connections and Accessories
When the pump is new, the cylinder usually unscrews without much
difficulty. Pumps that have been used for a while will be a little
harder to unscrew the cylinder. A strap wrench or the wrench kit
available from Teledyne ISCO can be used to assist with removing
the cylinder. Instructions for using this wrench package will be pro-
vided in Section 2.4.3 "Using the Wrench Package".
DO NOT USE A COMMON TOOL WITH TEETH (a monkey wrench
or water pump pliers, for example) TO REMOVE THE CYLINDER.
Doing so will damage the cylinder, resulting in leakage of the
temperature control fluid. Such damage is not repairable and will
require cylinder replacement.
11. If the wrench package was used to remove the cylinder, remove
the wrench package clamps before taking the cylinder off.
12. When the pump cylinder is free of the threads, hold the pump
cylinder with both hands and pull the cylinder straight up and
off the piston and push rod.
Some working fluid may spill out when the cylinder is removed. It
may be helpful to wrap an absorbent cloth around the base of the
cylinder to collect this fluid when the cylinder is pulled off.
As mentioned above, sometimes the cylinder is difficult to rotate, espe-
cially if salt buffers have accumulated in the cylinder threads. The
wrench package available from Teledyne ISCO will make this task
easier without damaging the cylinder.