Figure 4-3 Pump outlet connections
4.2.3 Dual-Gradient
System
Connections
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
Section 4 Gradient Pumping for Pressure, Flow, and Concentration Modes
spacer block
2-way
valve
Connection of the dual-gradient system requires two refill valve kits
and a gradient mixer.
The user supplied static mixer must have a pressure rating greater
than the maximum rated pressure of the pump. The static mixer can
be replaced by a user-supplied dynamic mixer.
Solvents are fed from each pump through the inline filters and check
valves, and into the static mixer, where they are mixed and fed into
your system apparatus. User supplied components are required. A
suggested mixing arrangement is shown in Figure 4-4 and Figure 4-5.
FROM REFILL
KIT ASSEMBLY
TO
APPARATUS
NOTE
pre-bent
SST tubing
5.1 cm length of
1/8" SST Tubing