3.9.2 Independent
Control of up to
Four Separate
Pumps
EAR99 Technology Subject to Restrictions Contained on the Cover Page
SyriXus Series Pumps Installation and Operation Guide
A SyriXus series syringe pump controller can run up to four syringe
pumps independently of each other in either Constant Pressure, Con-
stant Flow Mode, Dispense Mode, or any combination of the three. To
set up this option, use the following procedure.
1. Press
.
MENU
2. Press softkey 'A', MORE.
3. Press number '4', MULTI PUMP. The multi-pump menu will
appear.
4. Press number '1' for pumps A or B, press number '2' for pumps C
or D.
5. Press number '4', INDEPENDENT. The controller will set the
pumps to Independent Mode. Number 4 will blink, indicating
that Independent Mode is selected.
6. Select the HOLD PRESS or NORMAL mode of operation. Press
softkey 'A' to toggle between the two modes.
• HOLD PRESS
In Constant Pressure Mode, after the pump is empty, if the
outlet pressure rises past the setpoint, the pump will restart
and run the system down to the setpoint pressure. This
process will function until the cylinder is full.
• NORMAL
This feature shuts the system down if a pump runs empty in
Constant Pressure Mode.
Once the pumps have been set to this mode, they will operate
independently from one another. Each pump will operate at
its defined limit and rate. Independent Mode is the default
setting for the pump.
When you select a command such as
display will prompt you to designate which pump to stop or
refill. Only the designated pump will stop, while the other
pumps will continue to run.
In an emergency situation, toggling the On/Standby switch will
cause all pumps to immediately stop. Also, pressing the stop key on
the controller twice will cause all pumps to stop.
7. Return to the run screen by pressing 'D' three times. Then press
D (
). The display will show each pump's information
SELECT PUMP
and allow you to select any pump for programming changes.
Section 3 Basic Programming and Operation
STOP
NOTE
or
, the
REFILL
3-17