SyriXus Series Pumps Installation and Operation Guide
Section 2 Fluid System Connections and Accessories
2.3 Fluid Connection
Accessories

2.3.1 Manual Refill Kit

Kit Installation
EAR99 Technology Subject to Restrictions Contained on the Cover Page
2-8
Using nitrogen, blow out any liquid that remains inside the cylinder,
transducer port, and tubing. Using shop air pressure is not recom-
mended due to the possibility of compressor oil being present in the
air.
Liquids expelled by compressed gases may cause injury. Wear
eye protection. Certain liquids also may require other personal
protective equipment. Refer to the applicable Safety Data Sheet
(SDS) for more information.
The optional accessories discussed in this section are used to make
fluid connections from the pump(s) to another apparatus.
When making fluid connections that use ferrules, be sure to use the
ferrules provided in the kit. Push the tubing completely into the con-
nector and finger-tighten. Then tighten with a wrench to clamp the
ferrules onto the tubing.
The optional manual refill kit provides a high pressure, two-way valve
that connects to the pump inlet to a fluid reservoir. The kit contains all
tubing and hardware necessary for valve installation. Kit components
and connections are shown in Figure 2-6.
1. To attach the two-way valve to the pump housing, use the valve
spacer block and screws provided.
a. For the 500x pump, screw the male adapter into the inlet
port of the pump.
2. Connect the pre-bent stainless steel tubing from the valve to the
pump inlet. Use the nut and ferrule to connect the tubing at the
inlet and the valve fittings to connect the tubing at the valve.
3. Connect the PTFE refill tubing (with the filter) to the port of the
two-way valve, using the nuts and ferrules supplied.
When connecting to pressurized sources in supercritical fluid appli-
cations, use the stainless steel tubing without a filter. An inline filter
is contained in the CO
letin
TB08 CO2 Applications and Technical
RISK OF INJURY. THIS EQUIPMENT PRODUCES
HAZARDOUS
APPROPRIATE TUBING AND CONNECTIONS AS
NOTED IN THE MANUAL.
WARNING
NOTE
connection package (refer to Technical Bul-
2
Notes.
DANGER
PRESSURES.
PLEASE
UTILIZE