Controller; Lubrication; Gear Train Lubrication And Motor Drive Service - Teledyne SyriXus 65x Installation And Operation Manual

Syringe pumps
Table of Contents

Advertisement

SyriXus Series Pumps Installation and Operation Guide
Section 9 Pump Maintenance, Troubleshooting, and Servicing

9.8.2 Controller

9.9 Lubrication

Configuration 1
Figure 9-7 Gear train lubrication and motor drive service
EAR99 Technology Subject to Restrictions Contained on the Cover Page
9-10
The voltages in the controller unit between circuit common (TP16,
TP5, TP6) and the test points should be:
TP32
TP12
TP43
TP44
TP20
TP9
TP7
These voltages are displayed during the Analog Input routine in the
Diagnostics section of the firmware (Section 9.6.2 "ANALOG
INPUTS").
The pump is a precision engineered instrument that must be lubri-
cated every two years or every 6,000 strokes (whichever comes first) to
ensure proper service life. The pump has an easy-to-access lube wheel
that keeps the main gears lubricated during operation. Refer to Figure
9-7, configuration 1 or 2, depending on your pump motor type. For
your convenience, a lubrication kit (containing Never-seez
®
DUOLEC
1608 lubricants) is available from the Teledyne ISCO web
store (P/N 601244271).
Ball Screw
Thrust Bearing
Closeup of Lube Wheel and Worm
Gear at Each Location
5 VDC 0.002 VDC
8.1 VDC to +13 VDC
15 VDC 0.6 VDC
15 VDC  0.6 VDC
3.3 VDC ± 0.0 VDC
+2.5 VDC ± 0.0 VDC
+1.5 VDC ± 0.0 VDC
NOTE
(worm gear
unseen)
Configuration 2
®