Removing The Case Top; Controller Case Top Removal; Controller Case Top Screws (2 Of 4 Shown) - Tbd - Teledyne SyriXus 65x Installation And Operation Manual

Syringe pumps
Table of Contents

Advertisement

SyriXus Series Pumps Installation and Operation Guide
Section 9 Pump Maintenance, Troubleshooting, and Servicing
9.7 Removing the
Case Top
9.7.1 Controller Case
Top Removal
Figure 9-5 Controller case top screws (2 of 4 shown) - TBD
EAR99 Technology Subject to Restrictions Contained on the Cover Page
9-8
Run the test again. The screen should display:
Rec: [whatever is typed from PC]
Sent: ****** Serial Test #### *****
Press any key to exit
Anything typed on the computer keyboard should appear in the top
line on the controller display. The computer should continually display
the serial test shown in the second line of the controller display.
For some maintenance procedures, the case top of the controller or the
pump may need to be removed. Because interior access is easier with
the controller, this method of troubleshooting should be used
whenever possible.
RISK OF ELECTRIC SHOCK - DISCONNECT THE ELEC-
TRIC POWER BEFORE SERVICING. ONLY TRAINED SER-
VICE PERSONNEL MAY REMOVE THE CASE TOP.
RISQUE DE CHOC ÉLECTRIQUE. COUPER L´ALIMENTA-
TION AVANT LA RÉPARATION. L´USAGER NE DOIT PAS
DÉMONTER L´INSTRUMENT OU DÉRANGER LE
MÉCANISME DEDANS. ADRESSER LA REPARATION
SEULEMENT AUX TECHNICIENS COMPÉTENTS.
Troubleshooting for a number of issues can be done on the controller
main circuit board. Remove the four screws holding the case top in
place (two screws on each side). Lift the cover straight up and off.