Section 2 Fluid System Connections and Accessories
2.4.2 Removing the
Pump Cylinder
EAR99 Technology Subject to Restrictions Contained on the Cover Page
2-12
Table 2-4 Temp. Control Jacket Packages
Description
Temperature Control Jacket Packages for:
65x
260x
500x, 500xv, 1000x
O-rings (2) for Temperature Control Jackets:
65x
260x
500x
1000x
Installing a temperature control jacket on a pump or replacing the
O-rings will require the partial disassembly of the pump. The jacket
surrounds the pump cylinder and the pump cylinder functions as the
inside wall of the fluid containment. Normally, this involves removing
the pump cylinder and cylinder cap together as a unit.
1. Run the pump until the display shows CYLINDER EMPTY. It
may be necessary to turn OFF auto-refill so that the pump stops
when the cylinder is empty. Refer to Section 3.4.3 "Refill" for
instructions on how to turn auto-refill ON or OFF.
2. Disconnect the fluid fittings from the cylinder cap.
3. With the pump inlet and outlet ports open to the air, press
REFILL on the controller. The pump should stop and display
CYLINDER FULL when the piston is at the bottom of the cylin-
der.
4. Toggle the STANDBY switch on the controller, turning OFF the
controller display.
5. Toggle the POWER switch on the front of the pump to turn OFF
the pump.
6. Unplug the pump from the MAINS power.
7. Disconnect the TRANSDUCER cable from the back of the pump
and wrap the cable around the cylinder cap. On pumps with
external transducers, it is possible to disconnect the cable from
the transducer instead.
8. Remove the front cover from the pump tower (if necessary) and
locate the cylinder locking screw near the base of the cylinder on
the front side.
9. Using a 1/8" Allen wrench, loosen the set screw locking the cylin-
der about two turns. Do not remove this set screw.
10. Unscrew the cylinder from the cylinder mount. It will take sev-
eral revolutions of the cylinder to completely unthread it.