A [CR] must start the network. The controller is checking for the
presence of unit #1 but will get no response in 200 ms because
there is no unit 1.
Network Controller and Master Unit #0
1R 5D[CR]
Still no response.
Network Controller and Master Unit #0
1R 5D[CR]
Still no response, so unit 1 will be dropped from the poll.
Network Controller and Master Unit #0
2R 5C[CR]
Checks for unit 2 but will get no response in 200 ms because
there is no unit 2.
Network Controller and Master Unit #0
2R 5C[CR]
Still no response.
Network Controller and Master Unit #0
2R 5C[CR]
Still no response, so unit 2 will be dropped from the poll.
In this way units 3-5 will be checked and dropped from the poll.
Network Controller and Master Unit #0
6R 58[CR]
Check for presence of unit 6.
Unit 6
R 8E[CR]
Unit 6 responds.
Network Controller and Master Unit #0
7R 57[CR]
Since unit 7 does not exist, it will be dropped from the polling
scheme.
Network Controller and Master Unit #0
6R008IDENTIFY84[CR]
The master verifies the fact that unit 6 is a Model ___D. Refer to
Table 3-2 "D Series Controller to SyriXus Series Pump" for the
model name returned for each of the SyriXus series pump
models.
In this example, the master and the network controller are a
single unit. If they were separate units, the master would send
the inquiry to the network controller; then the network
controller would send the message to the slave unit the next
time it is polled. The slave would respond with the message to
the network controller. The next time the master is polled, the
message would be relayed.