Section 9: Specifications
9.5.5 SpO
Measurement/Display
2

9.6 TRENDS

9-6
Range:
Pulse Rate:
% Saturation:
Accuracy:
Pulse Rate:
SpO
:
2
SpO
accuracies are expressed as plus or minus "X" digits (saturation
2
percentage points) between saturations of 70–100%. This variation equals
plus or minus one standard deviation (1SD), which encompasses 68% of
the population. All accuracy specifications are based on testing the
subject monitor on healthy adult volunteers in induced hypoxia studies
across the specified range. Adult accuracy is determined with
Oxisensor ® II D-25 sensors.
Types:
Memory Storage:
Data interval:
Graphical Format:
Display range:
Vertical Scaling:
Heart rate:
NIBP:
SpO
:
2
Respiration
Rate:
Temperature
°C:
Temperature
°F:
Tabular Format:
Display interval:
20 BPM to 250 BPM
0 % to 100%
±3 BPM
70 % to 100%: ±2 digits
0 % to 69%: Unspecified
Graphical and Tabular
12 hours, nonvolatile
20 seconds: (Stored data point is the average over
20-second interval)
One graph per vital sign
2 hours, scrollable
50 BPM to 100 BPM, 0 BPM to 250 BPM,
25 BPM to 125 BPM
50 mmHg to 100 mmHg, 0 mmHg to 150 mmHg,
0 mmHg to 300 mmHg
0% to 100%, 60% to 100%, 80% to 100 %
0 BPM to 20 BPM, 0 BPM to 50 BPM,
0 BPM to 150 BPM
15º C to 45º C, 33º C to 41º C, 35º C to 39 ° C
55º F to 155º F, 91º to 107º F, 96º F to 102° F
One table for all variables
Six fields per row (time and 5 vital signs)
Per NIBP measurement, or 15 minutes for no NIBP,
or 20 seconds during alarm condition.