Front Case Disassembly Procedures - Mallinckrodt NELLCOR NPB-4000 Service Manual

Patient monitor
Table of Contents

Advertisement

Section 6: Disassembly Guide

6.5 FRONT CASE DISASSEMBLY PROCEDURES

6-6
This section describes the steps to separate the front and rear case assemblies,
and the items that may be removed/replaced on the front case assembly.
Step B1
To separate front and rear case assemblies:
a) Remove handle as indicated in step A1.
b) Use screwdriver to remove four screws fastening Rear Case Assembly to
Front Case Assembly. Retain for reassembly.
c) Separate two major case assemblies. There is enough cable and tubing
slack to permit the two assemblies to remain at approximately a 90-degree
angle to each other.
Main ribbon
cable assembly
Procedure
Illustration
Procedure
Illustration
Connector to
NIBP tubing
printer PCB
(when installed)