Section 6: Disassembly Guide
6-12
Step C2
To remove battery housing:
a) Remove battery as described in step C1. Carefully remove two foam
battery pads from battery housing.
b) If a printer is installed, remove it as described in step A3.
c) If a printer is not installed, remove printer blanking cover by slipping small
flat-blade screwdriver into one of the slots on the blanking cover. Use
screwdriver to gently depress snap-tab on inside of cover, while pulling
cover away from monitor. After first tab is released, repeat process on
other side of cover and remove cover.