SECTION 4: POWER-UP DEFAULTS MENU AND DIAGNOSTIC MODE

4. POWER-UP DEFAULTS MENU AND DIAGNOSTIC MODE

4.1 INTRODUCTION

4.2 POWER-UP DEFAULTS MENU

4.1 Introduction
4.2 Power-up Defaults Menu
4.3 Diagnostic Mode
This section discusses use of the Power-up Defaults Menu to configure power-on
default settings, and the Diagnostic Mode to obtain service-related information
about the monitor.
The purpose of the Power-up Defaults Menu (Table 4-1) is to allow the
authorized user to create a "power-up default" for each setting in the
NPB-4000/C. Power-up defaults are the settings in effect each time the
NPB-4000/C is powered on. Once the Power-up Defaults Menu is entered,
physiological monitoring is terminated. The screen layouts do not display any
information associated with normal monitoring operation.
Use the following procedure to configure the power-up default settings for the
NPB-4000/C monitor:
1.
Set NPB-4000/C to normal monitoring mode.
2.
Adjust each accessible setting on monitor as desired.
Note: Use the techniques described in the operator' s manual. Such settings
include alarm limits, choice of display type for the graphic frames, and
ECG lead select.
3.
Use knob to invoke Set-up Menu (choose the screwdriver icon found along
the bottom of the display).
4.
Select menu item "Enter Power-Up Defaults Menu".
Note: Once selected, a pop-up box will appear with the text "Enter 3-Digit
Passcode". Use the knob to enter the passcode, 2 1 5. This passcode is
set at the factory and may not be changed.
5.
Enter Passcode. The Power-up Defaults Menu will now be present. The
available menu items are explained in Table 4-1. Make changes to these
menu items as desired.
4-1