3.3.4.6 General Operation
3.3.4.6.1 Operation with a Human Subject
Note: The accuracy of NPB-4000/C temperature measurements is ±0.1ø C
(±0.2øF). In the procedure below, add the tolerance of the simulator to
the acceptable range of readings.
4.
Press On/Standby switch to turn monitor on.
5.
After normal power-up sequence, verify temperature reads 37ø ±0.1ø C
(98.6øF ±0.2øF if Fahrenheit is selected as temperature units.)
6.
Turn monitor off.
The following tests are an overall performance check of the system:
3.3.4.6.1
Operation with a Human Subject
3.3.4.6.2
Serial Interface Test
Patient monitoring involves connecting the monitor to a human subject for a
qualitative test.
1.
Connect EC-8 sensor extension cable to monitor.
2.
Connect Nellcor Durasensor oxygen transducer, model
DS-100A, to sensor extension cable.
3.
Clip DS-100A to subject as recommended in sensor directions for use.
4.
Connect a CE-10 ECG cable to NPB-4000/C.
5.
Connect ECG leads to cable.
6.
Connect ECG electrodes to leads.
7.
Apply ECG electrodes to subject according to leads and electrodes
directions for use.
8.
Connect SHBP-10 blood pressure hose to monitor.
9.
Apply appropriate SCBP series blood pressure cuff to subject according to
the cuff directions for use.
10. Press On/Standby switch to turn monitor on and verify monitor is operating.
11. Monitor should stabilize on subject' s physiological signal in about 15 to 30
seconds.
12. Verify saturation and heart and respiration rates are reasonable for subject.
13. Press NIBP Start/Stop switch on front panel of monitor.
Section 3: Performance Verification
3-15