Mallinckrodt NELLCOR NPB-4000 Service Manual page 86

Patient monitor
Table of Contents

Advertisement

Section 9: Specifications
9.5.2 Respiration Measurement/Display
9-4
Response to irregular
rhythm. 3.1.2.1(e)
Heart rate meter
response time.
3.1.2.1(f)
Time to alarm for
tachycardia.
3.1.2.1(g)
Pacemaker pulse
rejection. 3.1.4.1,
3.1.4.2
Technique:
Range:
Accuracy:
Leads:
Display Sweep
Speeds:
Lead Off Condition:
a) Ventricular bigeminy - 80 BPM
b) Slow alternating ventricular bigeminy - 60 BPM
c) Rapid alternating ventricular bigeminy -
120 BPM
d) Bi-directional systoles - 89 to 96 BPM
a) Change from 80 to 120 BPM: 4 to 6 sec
b) Change from 80 to 40 BPM: 6 to 7 sec
Waveform 4(a) Amplitude
0.5 mV
1 mV
2 mV
Waveform 4(b) Amplitude
1 mV
2 mV
4 mV
With the exceptions noted below, the monitor will
reject all pacemaker pulses having amplitudes of
±2 to ±700 mV and pulse widths from 0.1 to 2 ms,
with and without under/overshoot. Following are
pacer pulse conditions that the monitor will reject,
but are less than the maximum range listed in the
standard:
Pacer pulse width = 2
ms, double pulse,
without over/undershoot,
no QRS
Pacer pulse width = 2
ms, single and double
pulse, with 2 mV
over/undershoot, no QRS
Pacer pulse width = 0.1
ms, double pulse, with 2
mV over/undershoot,
ineffective pacing
Trans-thoracic impedance
3 breaths/min to 150 breaths/min
±3 breaths/min
RA to LA
6.25 mm/s, 12.5 mm/s, 25 mm/s
Detected and displayed
Avg. Time to Alarm
4 sec
4 sec
4 sec
Avg. Time to Alarm
3 sec
3 sec
3 sec
Pacer amplitude =
±300 mV
Pacer amplitude =
±300 mV
Pacer amplitude =
+300 mV, -450 mV

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nellcor npb-4000c

Table of Contents