Table 5-4: Error Code Categories - Mallinckrodt NELLCOR NPB-4000 Service Manual

Patient monitor
Table of Contents

Advertisement

Section 5: Troubleshooting
5.6.3 Switches/Knobs
5-6
4.
If Error Code screen indicates same error has occurred previously, take
monitor out of service and contact Mallinckrodt Technical Services or
your local Mallinckrodt representative for advice on remedial action.
5.
If Error Code screen indicates no previous occurrences of this error, monitor
may be returned to service.
As a reference, Table 5-4 lists the general categories for other error codes. The
error code categories are shown only in hexadecimal format:

Table 5-4: Error Code Categories

Code (hex)
500xxxx
internal user interface error
501xxxx
remote serial port error
502xxxx
date and time error
503xxxx
NIBP error
504xxxx
front end error
505xxxx
alarm error
506xxxx
audio error
507xxxx
recorder error
508xxxx
trend error
509xxxx
flash memory data error
50axxxx
SpO
50bxxxx
ECG error
50cxxxx
power down task error
50dxxxx
on board diagnostic error
50exxxx
power monitor error
50fxxxx
temperature measurement error
510xxxx
internal user interface error
511xxxx
error handling error
512xxxx
pSOS (operating system) initialization error
513xxxx
serial driver error
Table 5-5 lists recommended actions to address problems with the knob and
front panel switches.