SECTION 3: PERFORMANCE VERIFICATION

3. PERFORMANCE VERIFICATION

3.1 INTRODUCTION

3.2 EQUIPMENT NEEDED

3.1 Introduction
3.2 Equipment Needed
3.3 Performance Tests
3.4 Safety Tests
This section discusses the tests used to verify performance following repairs or
during routine maintenance. All tests can be performed without removing the
NPB-4000/C covers.
If the NPB-4000/C fails to perform as specified in any test, repairs must correct
the problem before the monitor is returned to the user.
Table 3-1 lists the equipment required for performance verification.

Table 3-1: Required Test Equipment

Equipment
Digital multimeter (DMM)
Sensor extension cable
Durasensor ® oxygen transducer
Oxisensor ® II oxygen transducer
ECG cable
ECG electrodes
ECG leads
NIBP hose
NIBP cuff
Pulse oximeter tester
ECG simulator
NIBP simulator
Respiration simulator
Temperature simulator
Safety analyzer
Stopwatch
Description
Fluke Model 87 or equivalent
EC-8
DS-100A
D-25
CE-10
standard
LE series
SHBP-10
SCBP series
Nellcor SRC-2
medSim 300 or equivalent
Bio-Tek " BP Pump" or equivalent
medSim 300 or equivalent
medSim 300 or equivalent
Bio-Tek 601 Pro or equivalent
Manual or electronic
3-1