Mallinckrodt NELLCOR NPB-4000 Service Manual page 87

Patient monitor
Table of Contents

Advertisement

9.5.3 NIBP (Noninvasive Blood Pressure) Measurement/Display
9.5.4 Temperature Measurement/Display
Note: Systolic and diastolic blood pressure measurements determined with
this device are equivalent to those obtained by a trained observer using
the cuff/stethoscope auscultation method, within the limits prescribed
by the American National Standard, Electronic or automated
sphygmomanometers.
Technique:
Measurement Modes:
Auto:
Manual:
STAT:
Cuff Pressure Display:
Blood Pressure
Measurement Range:
Systolic:
Mean Arterial
Pressure:
Diastolic:
Pulse Rate Range:
Blood Pressure
Accuracy:
Pulse Rate Accuracy:
Initial Cuff Inflation:
Subsequent Cuff
Inflation:
Standards:
Technique:
Range:
Accuracy:
Oscillometric
Automatic BP measurements at intervals of 1, 3, 5,
10, 15, 30, 60, and 90 minutes
Single measurement initiated by NIBP Start/Stop
switch
Series of consecutive measurements for 5 minutes
10mmHg to 300 mmHg
60 mmHg to 250 mmHg
30 mmHg to 235 mmHg
20 mmHg to 220 mmHg
20 BPM to 200 BPM
Mean error and standard deviation per
ANSI/AAMI SP10, 1992
Greater of ±2 BPM or ±2% of pulse rate value
180 mmHg (factory default), selectable 100 mmHg
to 280 mmHg, in 20 mmHg intervals
Previous SYS + 50 mmHg
Meets performance standards of
ANSI/AAMI SP10-1992
Thermistor probe (YSI 400 series compatible)
15° C to 45° C
±0.1° C
Section 9: Specifications
9-5