Section 4: Power-up Defaults Menu and Diagnostic Mode
4-2

Table 4-1: Power-up Defaults Menu

MENU ITEM
Accept Current Settings
as Power-Up Defaults?
Adult/Neonatal Mode
Permanent Audible Alarm
Silence
Alarm Suspend
Auto-Set Limits
CHOICES
" Yes"
If " Yes" is chosen, the
current NPB-4000/C
" No"
settings become the
power-up defaults.
Neonatal mode is not
available at this time; thus,
when this item is selected,
the following pop-up
message will appear:
" Selection not available" .
" Make Available"
If " Make Available" is
chosen, the caregiver may
" Deny Access"
permanently silence the
audible alarm for a
particular parameter via the
Alarm/Limits Menu. Some
institutions may wish to
prevent audible alarms from
being permanently silenced.
If so, " Deny Access" should
be selected.
" Make Available"
If " Make Available" is
chosen, the caregiver may
" Deny Access"
invoke the Alarm Suspend
Mode by pressing and
holding the Alarm Silence
switch for 2 seconds. Some
institutions may wish to
prevent Alarm Suspend
from being invoked. If so,
" Deny Access" should be
selected.
" Make Available"
If " Make Available" is
chosen, the caregiver may
" Deny Access"
invoke the Auto-Set Limits
function via the Alarm/Limits
Menu. Some institutions
may wish to prevent
Auto-Set Limits from being
invoked. If so, " Deny
Access" should be selected.
EXPLANATION