Section 16: MP-205 Service Manual; 044540A-0296
16.19.3 Repackaging in a Different Carton

16.20 MP205 SPECIFICATIONS

16.20.1 Measurement Range/Accuracy
16-14
If the original carton is not available:
1.
Place module in antistatic bag.
2.
Locate suitable corrugated cardboard shipping carton.
3.
Fill bottom of carton with packing material.
4.
Place bagged unit on layer of packing material and fill box completely with
packing material.
5.
Seal carton with packing tape.
6.
Label carton with shipping address, return address, and RGA number.
Range
Saturation:
Pulse Rate:
Accuracy
1
SpO
2
Adults:
Neonates:
Pulse Rate
1
Accuracies are expressed as ± "X" digits (saturation % points) between
saturations of 70 and 100 percent. This variation equals ± one standard
deviation (1SD), which encompasses 68 percent of the population. All accuracy
specifications are based on testing the subject monitor on healthy adult
volunteers in induced hypoxia studies across the specified range. Adult accuracy
is determined with Oxisensor
determined with Oxisensor II N-25 sensors. In addition, the neonatal accuracy
specification is adjusted to take into account the theoretical effect of fetal
hemoglobin in neonatal blood on oximetry measurements.
1 - 100%
20 - 250 beats per minute (bpm)
70 - 100%
0 - 69%
70 - 100%
20 - 250 bpm
®
II D-25 sensors. Neonatal accuracy is
±2 digits
unspecified
±3 digits
±3 bpm