Electrical; Environmental - Mallinckrodt NELLCOR NPB-4000 Service Manual

Patient monitor
Table of Contents

Advertisement

Section 9: Specifications

9.3 ELECTRICAL

9.4 ENVIRONMENTAL

9-2
Protection Class:
Degree of Protection:
Mode of operation:
Internal Battery:
Battery Operating Time:
NPB-4000
NPB-4000C
AC Mains:
DC (External):
Mechanical Shock:
Mechanical Vibration:
Sinusoidal:
Temperature:
Operating:
Storage:
Humidity:
Operating:
Storage:
Water Resistance:
Altitude:
Electromagnetic
Compatibility:
Class I, internally powered equipment:
per IEC 601-1, clause 2.2.4
Type CF: per IEC 601-1, clause 2.2.26
Continuous
6V, 8 Ampere-Hours; Type - sealed lead-acid
4 hours typical @ 25øC, no printing, one NIBP
measurement per 15 min. (fully charged battery)
3 hours typical @ 25øC, no printing, one NIBP
measurement per 15 min. (fully charged battery)
100 VAC to 240 VAC, 50 Hz to 60 Hz, 1A max
10 VDC to 16 VDC, 5A max
IEC 68-2-27
100g; 6 msec; three axes
18 total shocks, nonoperating
IEC 68-2-6
10 Hz to 58 Hz; 0.15 in. displacement
58 Hz to 150 Hz; 2g acceleration
4 min/sweep; 20 sweeps/axis, nonoperating
0 º C to 50øC
-20 º C to 60øC
5 % RH to 95% RH, noncondensing
5 % RH to 95% RH, noncondensing
IEC 529 Classification IPX1 (Protected against
vertically dripping water)
0 ft to10,000 ft (operating)
(1037 hPa to 697 hPa)
Radiated and conducted electromagnetic energy per
CISPR 11, Class B